| AccUView Wastewater UV %Transmission Analyzer
|
With over 30 years of optical design technology expertise HF scientific has developed it's 4th generation Wastewater UV%Transmission Analyzer specifically for the harsh environment of wastewater applications. The AccUView Wastewater is manufactured using advanced technology, a NEMA 4X Stainless Steel Enclosure and revolutionary ultrasonic technology making it the most reliable, maintenance free and cost efficient UV%T Analyzer for Wastewater. The Auto Clean Ultrasonic Cleaning System continuously cleans the optical chamber providing a maintenance free reliable measurement. |
| |
|
| Standard Features |
 |
| |
| Optical Design: |
| New optical design allows consistent readings with laboratory %T Analyzers |
| Bellows Pump: |
| Pump included to provide reliable sample flow to the instrument. |
| Certified Traceable Standards: |
| HF scientific inc certifies the PrimeTime 100%T Calibration Standard is prepared as outlined in Standard Methods for the Examination of Water and Wastewater, 21st Edition. Method 1080 A-C, Methods for Preparation of Reagent Water. |
| Backlight Display: |
| User adjustable Backlight Display allows viewing in low light conditions. Stainless Steel Enclosure includes large viewing window to allow monitoring of readings without opening the enclosure. |
| Certification: |
| Listed or Certified to CE, UL, CSA(ETL, ETLc) |
| Made in the USA |
|
| Related Products |
AccUView UV %T Monitor |
UVT-15 Transportable UV %T |
 |
 |
| |
|
| SPECIFICATIONS |
|
| Range |
0 - 100% Transmission ( Auto Ranging) |
| Resolution |
±0.1%T |
| Repeatability |
±0.1%T |
| Accuracy |
±1.0%T |
| Wavelength |
Ultra Violet 253.7nm |
| Response Time |
Updated at Selectable sample intervals from 10 to 60 minutes |
| Cleaning System |
Ultrasonic Cleaning |
| Outputs |
4-20 mA with isolator or Optional RS-485 with modbus protocol |
| Alarm |
User Settable 240 VAC form C Relay |
| Operating Temperature |
-20°C to 50°C (-4° to 122°F) Low temperature operation requires optional cold weather kit |
| Sample Temperature |
0°C to 40°C (32°F to 104°F) |
| Storage Temperature |
-20°C to 60°C (-4°F to 140°F) |
| Sample Lift |
3.5m (12ft) |
| Sample Strainers |
Foot Strainer and T-Strainer |
| Pump Type |
Bellows |
| Flow Rate |
Approximately 500ml/minute |
| Display |
Multiline Custom LCD with Backlight |
| Power Requirements |
100 to 240 VAC 47-63 Hz |
| Enclosure |
Stainless Steel with Lexan Window |
| NEMA Ratings |
Designed to meet NEMA 4X |
| Agency Testing |
CE, CE safety self declare, ETLus & ETLc |
| Dimensions |
20"W x 20"H x 13"D (50.8cm x 50.8cm x 33.2cm) |
| Shipping Weight |
22.6 kg (50lbs) |
|
|
| |
|
| |