SPECIFICATIONS
|
|
Range:
|
Auto
Ranging, 0-100 %T Transmission |
|
Resolution: |
±1.0%
of Reading |
|
Accuracy: |
±1% of
Full Scale |
| Wavelength: |
Ultra Violet 253.7nm |
| Response Time: |
Reading is Updated
every 30 seconds |
| Operating
Temperature: |
-4°F (-20°C) to 122°F (50°C) |
| Storage Temperature: |
-4°F (-20°C) to 140°F (60°C) |
| Optical Cleaning: |
Removable Cuvette
plus Inline Cleaning Solution Connection |
| Flow Rate: |
140-180ml / min |
|
Analyzer Separation: |
Up to
1000ft (300m) |
|
Communication: |
RS-485
Serial Port, 9600 Baud, Isolated 1.5kv |
|
Sensor/Sampler Enclosure: |
Stainless Steel |
|
Analyzer Case Dimensions: |
330 x
290 x 200mm (13" x 11.5" x 8") |
|
Calibration: |
Automatic, w/Calibration periods User Selectable |
|
Keyboard Entry Analyzer: |
8
Membrane Function Keys |
|
Built
In Diagnostics: |
Yes |
|
Analog
Output: |
4-20
mA and .1, 1 or 10 Vdc |
|
Alarms: |
1
System / Two Settable Hi, Lo or Off with 2.5. 5.0 or 10% Settable
Deadband |
|
Alarm
Contact Rating: |
Max.
250
VAC @ 5.0A Resistive, Isolated,
Form C (SPDT) |
|
Certification: |
CE,
ETL (UL), ETLC (CSA), NEMA 4X,IP66 |